Neonatal Baby Infant Radiant Warmer With LED Phototherapy
Neonatal Baby Infant Radiant Warmer With LED Phototherapy
Features
Microprocessor based servo controlled temperature system
The microprocessor-controlled unit allows for three modes: pre-warm, manual and automatic
APGAR timer to record therapy time
Audible and visual Alarm functions :
Power failure, temperature deviation, temperature sensor failure, skin temperature failure, over temperature
LED observation light adjustable,brightness adjustable
Heater head can be adjusted ±9 0°horizontally
Convenient X-ray cassette under infant bed
Tilt angle of infant bed is adjustable
Four Transparent protector folded with 2 infusion seals
Integrated low pressure aspirator
LED phototherapy unit
PVC resuscitator
With RS-232 connector,2 trays and 2 drawers
Technical data
| Power supply | AC110/220V 60/50Hz |
| Power input | 1100VA |
| Environment temperature | 18℃~30℃ |
| Environment relative humidity | 30%~75% |
| Atmospheric Pressure | 700~1060hpa |
| Environment air velocity of flow | <0.3m/s |
| Skin Temperature control range | 32℃~3 8℃ |
| Skin temperature control accuracy | ≤0.5℃ |
| Skin temperature sensor accuracy | ±0.3℃ |
| Mattress temperature uniformity | <2℃ |
| Infant Bed tilt angle | ±10° |
| LED observation light | 3W |
| Warm-up time(from 25℃) | <30min |
| Heater head horizontal | ±90° |
| Distance from heater to mattress | 80cm |
| Mattress size | 67cm*54cm |
| Timer range | 0~99h59min |
| LED phototherapy light 36cm | 620-2800μw/cm2 |
| LED phototherapy light 50cm | 400-1620μw/cm2 |
| Max vacuum adjustable | ≥-25kpa(-188mmHg) |
| Storage bottle | 1000mL |
Main body (including the Radiant source, Control system,Infant bed, Bracket), X-ray cassette, Observe light , I.V. pole, Skin temperature sensor, Tray, Mattress, Transparent protector, Castors. Drawers,PVC infant resuscitator, LED phototherapy unit, Low pressure aspirator,RS-232
Optional functions

Transport and storage
Each unit is packed one plywood case
Size:118*90*87cm
N.W.:80Kg G.W.:110Kg
Environment temperature -40℃~+55℃
Environment relative humidity ≤95%
Atmospheric Pressure 500~1060hpa