High-strength Anesthesia Machine with Ventilator

MCS-A03
minimum order
1
Stock Time
3-15 working daysDay
Quantity:
Contact Now
add to cart
All Products

High-strength Anesthesia Machine with Ventilator

AnesthesiaMachine

Main body

High-strength engineering plastic rack. light, beautiful and corrosion resistant

Scope of application

adult, Child

Gas source

O2: 0.28-0.6MPa     N2O: 0.28-0.6MPa

Rapid oxygen flow

O2:0.1-1.0L/min     1.1-10L/min

N2O: 0.1-1.0L/min   1.1-10L/min

O2, N2O linkage and N2O Stopper

When using nitrous oxide, oxygen concentration>25%;

When the oxygen pressure<0.2MPa. the flow of nitrous oxide would be cut off.

Flow rate d rapid oxygen supply

25-75L/min

Vaporizer

it has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0~5vol%. Among Halothane, Enflurane, Isoflurane and Sevoflurane, two can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available.

Low oxygen pressure alarm

There will be sound alarm when the oxygen pressure < 0.2MPa

Respiratory circuit

work mode: all close, semi-close, semi open

APL: ≤12.5kPa

Respiratory Bellows

bellows for adults, bellows for children tidal volume range:0~1500ml

 Ventilator

Display mode

High-definition 5.7' LCD screen display

Ventilation mode

IPPV, SIPPV, SIMV, MANUAL

Ventilation function

PEEP, SIGH, IRV

Ventilation parameters

Tidal volume

50-1500ml

Rate

2-99bpm

SIMV rate

2-20bpm

I:E

2:1-1:8

Inspiratory trigger pressure

-10~20cmH2O

PEEP

02~20cmH2O

Pressure range

10~60cmH2O

SIGH

1.5 times the inspiratory time/100bpm

Parameter for ventilation monitoring

tidal volume. Minute Ventilation volume. IPPV rate. SIMV rate. total respiratory rate. I/E. peak pressure of airway. pressure -time waveform. flow rate -fire waveform. PEEP. inspiratory trigger pressure.( average pressure, oxygen concentration. battery capacity)

Security Alert System

Oxygen concentration alarm

upper limit setting range

low limit setting range

(50%-99%)

(15%-80%)

Airway pressure alarm

upper limit setting range

low limit setting range

5-60cmH2O

0-50cmH2O

minule ventlation volume alarm

upper limit setting range

low limit setting range

1-30L/min

0-29L/min

tidal volume

upper limit setting range

low limit setting range

10-2000ml

0-1800ml

sustained high-pressure alarm

it will give alarm when stress have consistently been higher than 25cmH2O

(Suffocation alarm, Power alarm, Intubation off alarm, battery voltage alarm)

Power

AC 100-240V 50/60Hz(UPS. Storage battery)

(Mechanical Arms)

(End-tidal CO2 monitoring)

(Specific interface of waste anesthesia gas)

Items within brackets are optional.